• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Inspected 1 opened/used sensor and performed visual inspection and found cannula separated from tubing.Unable to confirm customer received sensor in said condition due to customer returned opened/used.
 
Event or Problem Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 110 mg/dl and the sensor glucose was 50 mg/dl at the time of the incident.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7964192
Report Number2032227-2018-42594
Device Sequence Number221782
Product Code OZO
UDI-Device Identifier20643169646432
UDI-Public(01)20643169646432(017)20180313
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/13/2018
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG22011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2018
Initial Report FDA Received Date10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight163
-
-