• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551PNAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551PNAP
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 10/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had compromised force sensor system alarm.The customer blood glucose level was 59 mg/dl, 55 mg/dl, 90 mg/dl,185 mg /dl 125 mg/dl and 72 mg/dl.The customer was assisted with troubleshooting.Customer stated that they took battery out the insulin pump, screen was blank but insulin pump was still alarming.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump received compromised force sensor system alarm during the basic occlusion test due to loose/protruded or flush drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test or test for motor error alarm due to compromised force sensor system alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551PNAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7964574
Report Number2032227-2018-42866
Device Sequence Number170225
Product Code OZO
UDI-Device Identifier00643169507296
UDI-Public(01)00643169507296
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551PNAP
Device Catalogue NumberMMT-551PNAP
Device Lot NumberA3551PNAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Initial Date Received by Manufacturer 10/02/2018
Supplement Date Received by Manufacturer01/10/2019
Initial Report FDA Received Date10/15/2018
Supplement Report FDA Received Date01/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight185
-
-