| Catalog Number |
HLF-S365-HSMA |
| Medical Device Problem Codes |
Break (1069); Arcing (2583)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Code Available (3191)
|
| Date of Event |
09/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # - k124030.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
It was reported during a ureteroscopy with laser lithotripsy , the cook® single-use holmium laser fiber broke at the entrance of the scope.The urology resident was handling the scope and the fiber broke at the entrance of the scope and shocked the resident and left a mark on her wrist.The holmium laser was shut off immediately and there was no harm to the patient nor was there anything left inside the patient.Everything happened outside the patient at the scope but the resident did have a mark on her wrist from where the fiber broke and touched her wrist.Additional information provided on 03oct2018.It was reported that the resident received a small red mark on her wrist and one day later she said she was fine.The resident did not receive medical treatment due to this issue.As reported, the resident was manipulating the ureteroscope while the laser fiber was in the working channel of the scope.She turned her wrist to manipulate the scope and went too far.As there was not enough slack on the laser fiber, it broke and left a mark on her wrist.The laser was shut off promptly.The case was continued and performed as scheduled.There were no additional procedures required as a result of this event and there were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint device was not returned for an evaluation and no photograph or imaging was provided.Without the complaint device, a device failure analysis was unable to be performed.A document based investigation was conducted including a review of complaint history, the device history record, instructions for use, manufacturing instructions, quality control data, specifications, and trends.A review of the device history did not observe any non-conformances that may have contributed to this incident.A review of complaint history records revealed no other complaints associated with the complaint device lot number 8772231.This device is shipped with the instructions for use (ifu) which contains the following precautions: ¿ never subject fiber optics to sharp bends in handling, use, or storage.¿ discard any fiberoptic assembly that is cracked or broken, or does not meet minimum transmission standards.¿ do not exceed recommended power limits.The device is inspected visually and functionally for cracks and fractures by manufacturing and quality control and no notable gaps in production or processing controls were noted.Based on the provided information the investigation conclusion is likely the result of unintentional use error.Per the quality engineering risk assessment, no additional risk mitigating activity is required at this time.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
There has been no new information received since the last report was submitted.
|
| |
|