• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH ACCUMAXTM SINGLE USE HOLMIUM LASER FIBER; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH ACCUMAXTM SINGLE USE HOLMIUM LASER FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M0068404030
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exact patient age is unknown; however the patient was reported to be over 18 years old.(b)(4).Investigation results: one accumax 550 laser fiber was received for evaluation.Visual examination revealed that the exposed glass fiber tip measured approximately 3.5 mm and appeared unused.Additional examination under magnification revealed that the pattern on the tip face was consistent with a slight degradation.There was no damage along the body of the fiber.Visual examination of the fiber face within the sma connector under magnification revealed that there was debris on the fiber face and that only a small amount of light was coming through.The condition of the returned device would cause the device to emit insufficient energy transmission during ablation thereby confirming the event.The noted damages indicate difficulty was experienced during the procedure and was likely due to anatomical or procedural factors such as tortuous anatomy or maneuvering of the device.Therefore, a review and analysis of all available information indicated that the most probable root cause is operational context.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 550 laser fiber was used during ureterolithotripsy procedure in the ureter performed on (b)(6) 2018.According to the complainant, during the procedure and outside the patient, according to the complainant, during the procedure and outside the patient, it was noted that the tip of the laser fiber burnt out while it was connected with the laser unit.The procedure was completed with another accumax 550 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed that there was debris on the fiber face within the sma connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUMAXTM SINGLE USE HOLMIUM LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
MDR Report Key7964983
Report Number3005099803-2018-02611
Device Sequence Number1395400
Product Code GEX
UDI-Device Identifier08714729784609
UDI-Public08714729784609
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K082928
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2020
Device Model NumberM0068404030
Device Catalogue Number840-403
Device Lot NumberC000004728
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2018
Initial Report FDA Received Date10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-