It was reported a turbo-ject® double lumen power-injectable picc placed on (b)(6) 2018 in a patient's arm was found to have blood backing up in the tubing.It was also noted tpn was leaking from the purple hub which was observed to be cracked.The line was rewired with a like device utilizing the same insertion site.Additional information provided by the facility stated that if needleless connectors become too tight, forceps are used over gauze to disconnect the pair.The device was discarded by the facility and will not be returned for investigation.As reported, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation - evaluation: a review of the complaint history, device history record, instructions for use (ifu), and quality control was conducted during the investigation.The complaint device was not returned; therefore, no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Based on the information provided, no product returned, and the results of our investigation, a definitive root cause could not be established.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|