• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER Back to Search Results
Catalog Number UPICDS-4.0-CT-NT-1111
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a turbo-ject® double lumen power-injectable picc placed on (b)(6) 2018 in a patient's arm was found to have blood backing up in the tubing.It was also noted tpn was leaking from the purple hub which was observed to be cracked.The line was rewired with a like device utilizing the same insertion site.Additional information provided by the facility stated that if needleless connectors become too tight, forceps are used over gauze to disconnect the pair.The device was discarded by the facility and will not be returned for investigation.As reported, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, device history record, instructions for use (ifu), and quality control was conducted during the investigation.The complaint device was not returned; therefore, no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Based on the information provided, no product returned, and the results of our investigation, a definitive root cause could not be established.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-JECT® DOUBLE LUMEN POWER-INJECTABLE PICC
Common Device Name
LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7965052
Report Number1820334-2018-03059
Device Sequence Number1706516
Product Code LJS
UDI-Device Identifier00827002345140
UDI-Public(01)00827002345140(17)200109(10)8499685
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K132334
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/09/2020
Device Catalogue NumberUPICDS-4.0-CT-NT-1111
Device Lot Number8499685
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/24/2018
Supplement Date Received by Manufacturer10/24/2018
Initial Report FDA Received Date10/15/2018
Supplement Report FDA Received Date11/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age17 MO
Patient Weight13
-
-