RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Electrical /Electronic Property Problem (1198); Connection Problem (2900)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2018.Date of report: 15oct2018.Philip's field service engineer (fse) evaluated the device and found no damage to the alternating current inlet connector.The ventilator was operating "fine", however, the device found to be operating on internal battery only.The fse replaced the power management (pm) printed circuit board assembly (pcba) to address the issue.The ventilator passed all testing and operated within the manufacturing specifications.
|
|
Event Description
|
The customer reported that the ventilator needed the alternating current (ac) connector replaced.The ventilator was not in use on a patient at the time of the event.The event date was not specified; estimate used.
|
|