| Model Number |
04.001.002 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Patient code (b)(4) used to capture additional medical/surgical intervention required.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, during implantation of a titanium humeral nail ex, the surgeon decided to use a titanium end cap and placed in the proximal tip of the nail.During the implantation procedure, the end cap would not advance to the full sitting.When the surgeon decided to remove it and re-implant the end cap, the implant would not come out.Efforts were made with several instruments and the aid of t25 screw driver but the location and space within the humeral head made it impossible to extract the end cap.After an hour or so of efforts, the surgeon decided to stop the extraction and leave the implanted end cap in the patient.The end cap is about 2-3mm proud at the superior aspect of the humeral head.There was a surgical delay of one hour or more.The procedure was completed successfully.The patient outcome is unknown.Concomitant device reported: unknown titanium humeral nail ex (part # unknown, lot # unknown, quantity 1).This report is for one (1) ti end cap t25 strdrv 10mm ext humeral nail-ex spiral blade.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The nail being implanted was a titanium (ti) humeral nail ex 7mm x 220mm.Concomitant devices: ti humeral nail ex (part: unknown, lot: unknown, quantity: 1), t25 screwdriver (part: unknown, lot: unknown, quantity: 1).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is not expected to be returned for manufacturer review/investigation.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The investigation was performed based upon a review of the received medical images as no product was received.Upon review of the medical images, it can be confirmed that the end cap is not fully inserted into the nail.The cause for the issue is not able to be determined.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|