One 72202754 biosure sync tensioner reported on.This is a four year old reusable device.Product was not available for evaluation.Definitive conclusions, accurate investigation and evaluation were not fully possible without evaluation of physical product.The complaint could not be ultimately confirmed.If objective evidence, relevant information, packaging or product becomes available to assist with evaluation, the complaint will certainly be revisited.Factors that may affect device performance include: device ability, product knowledge.Influences that could compromise product performance or integrity that are unrelated to manufacture include: compliance with recommended maintenance and care for the multiple components of the product.Repeat handling during disassembly, cleaning, sterilization reassembly.Use of incompatible cleaning agents that do not meet ph requirements for the product materials.Use of force before, during or after surgical procedure.This product is disassembled, cleaned and sterilized repeatedly.Inspect the device for wear or cracking.Discard the device if any wear is observed or if the tip becomes dull.Pay careful attention to cleaning devices with challenging design features.Challenging design features can include, but are not limited to, suction levers, stopcocks, interfaces, cannulations, holes, blind holes, crevices, hinges, mating surfaces, etc.Low-sudsing, neutral 6.0¿8.0 ph, enzymatic detergents are recommended.Do not use detergents above 11.0 ph.Use deionized water for washing and rinsing unless otherwise specified.As with any surgical instrument, careful attention should be exercised to ensure that excessive force is not placed on the instrument.Excessive force can result in instrument failure.This product met predetermined specifications upon release to distribution.No further investigation is warranted at this time.
|