BOSTON SCIENTIFIC CORPORATION WALLFLEX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00570540
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex biliary rx covered stent was to be used to treat a malignant stricture in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2018.Reportedly the patient's anatomy was tight.According to the complainant, at the end of the stent placement procedure, angiography was performed and noticed that the stent moved out from the original location.The stent was removed with forceps.The procedure was completed with another wallflex biliary stent.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).Results: a fully deployed wallflex biliary stent was received for analysis; the delivery system was not returned.Visual analysis of the returned device found that no issues noted with the device.The patient's anatomy was reported to be tight, which may have contributed to the issue with stent positioning.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.Additionally, "stent migration" and "stent misplacement" are noted within the dfu as potential adverse events associated with the use of this device.Therefore, the most probable root cause is known inherent risk of device.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a wallflex biliary rx covered stent was to be used to treat a malignant stricture in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2018.Reportedly the patient's anatomy was tight.According to the complainant, at the end of the stent placement procedure, angiography was performed and noticed that the stent moved out from the original location.The stent was removed with forceps.The procedure was completed with another wallflex biliary stent.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key7965770
Report Number3005099803-2018-60848
Device Sequence Number19259171
Product Code FGE
UDI-Device Identifier08714729764922
UDI-Public08714729764922
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K140630
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/22/2019
Device Model NumberM00570540
Device Catalogue Number7054
Device Lot Number0021167828
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Initial Date Received by Manufacturer 09/19/2018
Supplement Date Received by Manufacturer11/07/2018
Initial Report FDA Received Date10/15/2018
Supplement Report FDA Received Date12/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient Weight150
-
-