• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1.5MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1.5MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035421520
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Subject device is not available.
 
Event or Problem Description
During coil embolization of aneurysm located at the middle cerebral artery (mca) bifurcation, the coil was repositioned 2-3 times.The coil detached and confirmed through fluoroscopy.It was observed that 0.5cm of the coil was hanging out of the neck of the aneurysm and was found to be stretched.A stent was deployed in order to prevent the coil hanging in the vessel.There were no clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly, and performance specifications.Only the delivery wire was returned as the main coil remains implanted.No functional test could be performed on the returned device.The review and analysis of all available information failed to indicate an assignable cause.
 
Event or Problem Description
During coil embolization of aneurysm located at the middle cerebral artery (mca) bifurcation, the coil was repositioned 2-3 times.The coil detached and confirmed through fluoroscopy.It was observed that 0.5cm of the coil was hanging out of the neck of the aneurysm and was found to be stretched.A stent was deployed in order to prevent the coil hanging in the vessel.There were no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1.5MM X 2CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7967620
Report Number3008881809-2018-00440
Device Sequence Number13524313
Product Code HCG
UDI-Device Identifier04546540697981
UDI-Public04546540697981
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Catalogue NumberM0035421520
Device Lot Number20264539
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Initial Date Received by Manufacturer 10/08/2018
Supplement Date Received by Manufacturer12/13/2018
Initial Report FDA Received Date10/15/2018
Supplement Report FDA Received Date12/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
-
-