| Model Number |
760 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the service engineer evaluated the ventilator and replaced the pressure solenoid (psol) printed circuit board (pcb); after repairs the unit passed all tests and manufacturer¿s specifications.The replaced psol pcb was returned to medtronic for analysis.The analysis isolated the fault to electronic component u37.The reported complaint mode will be utilized for trending purposes and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the 700 series ventilator has failed the power on self test (post) and generated power on alarm.The ventilator was not in use on a patient at the time of the reported event.
|
| |
|