Model Number 760
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: the service engineer evaluated the ventilator and replaced the pressure solenoid (psol) printed circuit board (pcb); after repairs the unit passed all tests and manufacturer¿s specifications.The replaced psol pcb was returned to medtronic for analysis.The analysis isolated the fault to electronic component u37.The reported complaint mode will be utilized for trending purposes and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the 700 series ventilator has failed the power on self test (post) and generated power on alarm.The ventilator was not in use on a patient at the time of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
760 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway,gw
MDR Report Key7967628
Report Number8020893-2018-00503
Device Sequence Number10804406
Product Code CBK
UDI-Device Identifier10884521180932
UDI-Public10884521180932
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K990897
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number760
Device Catalogue NumberG-760220DIUA-EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2018
Initial Report FDA Received Date10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-