Model Number 9733560XOM |
Device Problems
Failure to Power Up (1476); Environmental Compatibility Problem (2929)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that when booting the system, it would not power on.User wiggled the cord and can sometimes get the green led power switch to illuminate briefly.There was no patient present.
|
|
Manufacturer Narrative
|
No parts have been returned.A medtronic representative went to the site to test the equipment.Testing revealed that system performed as intended.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
|