ABBOTT IRELAND ARCHITECT HBSAG QUALITATIVE CONFIRMATORY; HBSAG CONFIRMATORY Back to Search Results
Catalog Number 04P54-25
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.An evaluation is in process.
 
Event or Problem Description
The customer observed falsely (b)(6) results while using architect hbsag qualitative confirmatory reagents.The following data was provided.Sid 51403339 initial 0.68 s/co (b)(6), repeat 1.00 s/co (b)(6), 2742.65 s/co (b)(6) multiple hbsag results were (b)(6) while using the qualitative assay (2876.73 s/co, 2783.34 s/co, 2884.78 s/co).No impact to patient management was reported.
 
Additional Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, and a review of labeling.No adverse trend was identified for the customer's issue.No return patient sample was available.Historical performance of the reagent lot was evaluated using world wide data.The patient data was analyzed and within established control limits.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT HBSAG QUALITATIVE CONFIRMATORY
Common Device Name
HBSAG CONFIRMATORY
Manufacturer (Section D)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI  NA
MDR Report Key7967774
Report Number3008344661-2018-00088
Device Sequence Number16492691
Product Code LOM
UDI-Device Identifier00380740007799
UDI-Public00380740007799
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P110029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/13/2018
Device Catalogue Number04P54-25
Device Lot Number85274FN00
Initial Date Received by Manufacturer 09/20/2018
Supplement Date Received by Manufacturer03/05/2019
Initial Report FDA Received Date10/15/2018
Supplement Report FDA Received Date03/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT I2000SR LN 03M74-02; ARCHITECT I2000SR LN 03M74-02; LN 04P53-35 LN 88138FN00; LN 04P53-35 LN 88138FN00; SN ISR53423, ARCH HBSAG QUAL; SN ISR53423, ARCH QUAL
Patient Age59 YR
-
-