| Model Number |
15 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control replaced the interface pcb assembly and then completed other, unrelated, repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device would not power on when prompted.There was no patient use associated with the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device would not power on when prompted.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control further evaluated the removed interface pcb assembly and determined that the cause of the reported issue was current leakage between pins 23 and 25 of pcb-mounted jack connector, designator j31.
|
| |
|