| Model Number |
C42007E |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer reported that the surgery department is having issues with the tubing separating, falling on the floor and leaking after being connected to a patient.There was no patient harm.
|
| |
|
Event or Problem Description
|
|
The customer reported that the surgery department is having issues with the tubing separating, falling on the floor and leaking after being connected to a patient.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: removed incorrect lot number.The customer's report that the tubing separated was confirmed.Visual inspection showed a separation at the engagement between the set's first male exit luer end and female luer lock.After manually reconnecting the separated components, functional testing and pressure testing resulted in no further separations.The root cause of the customer's report could not identified.
|
| |
|
Event or Problem Description
|
|
The customer reported that the surgery department is having issues with the tubing separating, falling on the floor and leaking after being connected to a patient.There was no patient harm.
|
| |
|