• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION TECNIS CL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted, give date: not applicable, lens not implanted.If explanted, give date: not applicable, lens not implanted, therefore, not explanted.Attempts have been made to obtain missing information; however, to date no further information was provided.All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that a model z9002 intraocular lens had debris on it.This was observed during handling, but prior to insertion.No further information provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 10/19/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The lens was observed under a microscope.The lens was covered with what appeared to be blood, marks, and viscoelastic in the optic zone with one haptic distorted.The other haptic was observed positioned.Due the condition of the sample received, the complaint could not be verified since the debris could not be identified.No product quality deficiency was identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION
santa ana CA
MDR Report Key7968118
Report Number2648035-2018-01323
Device Sequence Number16587901
Product Code HQL
UDI-Device Identifier05050474530584
UDI-Public(01)05050474530584(17)220822
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2022
Device Model NumberZ9002
Device Catalogue NumberZ900200145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer11/06/2018
10/25/2020
Initial Report FDA Received Date10/15/2018
Supplement Report FDA Received Date11/13/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-