• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JMS SINGAPORE PTE LTD WINGEATER A.V.FISTULA 15GX1" BE 30CM W/CLAMP; JMS WINGEATER A.V.FISTULA NEEDLE SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JMS SINGAPORE PTE LTD WINGEATER A.V.FISTULA 15GX1" BE 30CM W/CLAMP; JMS WINGEATER A.V.FISTULA NEEDLE SET Back to Search Results
Model Number 820-5002-33
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiopulmonary Arrest (1765); Hemorrhage/Blood Loss/Bleeding (1888); Hypovolemia (2243)
Date of Event 09/21/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
Exemption number: e2010016.Jmss (manufacturer) is submitting the report on behalf of jmsna (importer).Our manufacturing process was within control and no abnormality was found on the given product lot.The product lot met all the qa outgoing inspection criteria prior releasing to the market.Based on the results of preserved samples evaluation and dhr analysis, there was no abnormality found on complaint lot.In addition, from the information gathered from (b)(4), user facility said that their suspicion might be due to patient movement.Furthermore, the user facility had emphasized there were no product problem in use as standard fmc taping policy was adhered.Hence, this complaint is not related to our manufacturing process.(b)(4).
 
Event or Problem Description
On (b)(6) 2018, a hemodialysis patient was going through a 3.5 hour hemodialysis treatment when the patient's arm was noted to have dropped onto chair, causing tension on venous line.The venous fistula needle was noted to be dislodged.Pressure was held to venous site; blood was normal and normal saline was administered via the arterial fistula needle.Oxygen was applied via nasal cannula, 911 was called.Patient did not arouse with painful stimuli; cpr was initiated with one round administered.Approximately 800ml of normal saline administered.Hemostasis achieved on venous cannulation site.The patient was then transported to er via ems and was transfused with 2 units packed red blood cells.Upon inspection of venous needle, tape and gauze dressing were still intact on the fistula needle.The patient was discharged from the hospital on (b)(6) 2018 but sustained a cardiopulmonary arrest in her home on (b)(6) 2018 which resulted in her death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WINGEATER A.V.FISTULA 15GX1" BE 30CM W/CLAMP
Common Device Name
JMS WINGEATER A.V.FISTULA NEEDLE SET
Manufacturer (Section D)
JMS SINGAPORE PTE LTD
440 ang mo kio industrial pk 1
singapore, 56962 0
SN  569620
MDR Report Key7968307
Report Number3002807350-2018-00007
Device Sequence Number10813509
Product Code FIE
UDI-Device Identifier08888483005291
UDI-Public08888483005291
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K111948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/12/2018
Report Date (Section F) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2023
Device Model Number820-5002-33
Device Lot Number180116351
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/01/2018
Device Age9 MO
Event Location Outpatient Treatment Facility
Date Report to Manufacturer09/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2018
Initial Report FDA Received Date10/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Hospitalization; Life Threatening;
Patient Age57 YR
Patient Weight52
Date Report Sent to FDA10/04/2018
-
-