| Model Number |
PED-450-30 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the pipeline failed to open in the middle section.More than 50% of the device had been deployed when it failed to open, with it resheated less than or equal to 2 times.As a result the pipeline was removed with the microcatheter.The patient was undergoing surgery to treat a ruptured aneurysm in the right internal carotid artery with a max diameter of 13mm and a neck diameter of 7mm.It was noted that the patient's vessel tortuosity was minimal and that the pipeline was not positioned in a bend.Dual antiplatelet treatment was administered.Ancillary devices includes a 9f cook sheath, marksman, echelon microcatheter, x-pedion guidewire, and axium coils.The devices were prepared as indicated per the ifu.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
As a result of the issue the stent was replaced.The cause of the issue was unknown.The devices were prepared/flushed as indicated per the ifu.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation the pipeline flex braid was returned for evaluation without its pushwire.The pipeline flex braid was returned detached from the pus hwire; therefore, the distal and proximal ends of the braid were unable to be identified.The pipeline flex braid appeared to be fully opened and moderately frayed at both ends.In addition, the middle section of the braid fully opened and no damage.No other anomalies were observed.Based on the analysis findings, the customer report of ¿failure to open at the middle section¿ could not be confirmed.The event cause could not be determined as the returned pipeline flex braid fully opened and moderately frayed at both ends.Additionally, the middle segment of the pipeline flex braid fully opened with no damage.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|