| Model Number |
E1450-6 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a thoracic procedure, they used the diathermy pencil and it was set to 50 with the medium tip and rubber shod to touch the forceps with it.They then noticed that the diathermy pencil has a flame on its tip and it was dropped on the floor.They stepped on the diathermy pencil tip that was on the floor and they used a new diathermy pencil with new medium tip and rubber shod and the setting was changed down to 20 to complete the case.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned electrode did not meet specification as received.Visual inspection found deformation and damage on the rubber shod and a small amount of charring on the electrode tip.The reported condition could not be confirmed, but the electrode showed damage consistent with heat damage.Inspection noted that the electrode tip was charred, and the coating was beginning to wear off.The rubber shod showed signs similar to thermal damage.The device was correctly assembled.Activating the device can result in arcing if other conductive objects are nearby, particularly at a higher power setting.Combustion can occur under these circumstances particularly when an enriched oxygen or nitrous oxide environment is present.The investigation identified the probable root cause of the reported event to be the user activating the device in such that the device arced.The instructions for use (ifu) states, both oxygen and nitrous oxide support combustion.Watch for enriched oxygen and nitrous oxide atmospheres near the surgical site, especially during head and neck surgery.Enriched oxygen atmospheres may result in fires and burns to patients or surgical personnel.Do not exceed the maximum power limits as stated in the ifu.Exceeding these power limits may result in patient injury or product damage.If the electrode is damaged, discard it.Always use the lowest power setting to achieve the desired surgical effect.The maximum power settings are detailed in the instructions for use.Some surgeons may elect to buzz the hemostats during a surgical procedure.This practice is not recommended and the hazards of such a practice probably cannot be eliminated.To minimize the risk, when using coated electrodes, place the edge of the electrode against the hemostat or metal instrument to assure a good electrical pathway.Do not activate the generator until the accessory makes contact with the instrument.Additional measures to minimize the risk of performing this technique are listed in the electrosurgical generator users manuals.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
New information for follow up (b)(4) was notified to manufacturer on 2018-11-12.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned electrode did not meet specification as received.Visual inspection found deformation and damage on the rubber shod and a small amount of charring on the electrode tip.The reported condition could not be confirmed, but the electrode showed damage consistent with heat damage.Inspection noted that the electrode tip was charred, and the coating was beginning to wear off.The rubber shod showed signs similar to thermal damage.The device was correctly assembled.Activating the device can result in arcing if other conductive objects are nearby, particularly at a higher power setting.Combustion can occur under these circumstances particularly when an enriched oxygen or nitrous oxide environment is present.The investigation identified the probable root cause of the reported event to be the user activating the device in such that the device arced.The instructions for use (ifu) states, both oxygen and nitrous oxide support combustion.Watch for enriched oxygen and nitrous oxide atmospheres near the surgical site, especially during head and neck surgery.Enriched oxygen atmospheres may result in fires and burns to patients or surgical personnel.Do not exceed the maximum power limits as stated in the ifu.Exceeding these power limits may result in patient injury or product damage.If the electrode is damaged, discard it.Always use the lowest power setting to achieve the desired surgical effect.The maximum power settings are detailed in the instructions for use.Some surgeons may elect to buzz the hemostats during a surgical procedure.This practice is not recommended and the hazards of such a practice probably cannot be eliminated.To minimize the risk, when using coated electrodes, place the edge of the electrode against the hemostat or metal instrument to assure a good electrical pathway.Do not activate the generator until the accessory makes contact with the instrument.Additional measures to minimize the risk of performing this technique are listed in the electrosurgical generator users manuals.If information is provided in the future, a supplemental report will be issued.
|
| |
|