Catalog Number 2N3326
Medical Device Problem Codes Fluid/Blood Leak (1250); Tear, Rip or Hole in Device Packaging (2385); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event occurred on an unspecified date in (b)(6) 2018.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the packaging was damaged and the tubing of an interlink system t-connector extension set was cut and sticking out of the packaging through the side seal.This was noted prior to use.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph revealed that a part of the set was trapped in the seal area.The reported condition was verified.The cause of the condition was determined to be a manufacturing issue.A batch review was conducted and it was found that during manufacturing, blisters with sets cut or trapped in the seal area of the blister during the packaging process were identified and discarded.Awareness training was provided to involved personnel.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SYSTEM T-CONNECTOR EXTENSION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - AIBONITO
aibonito
MDR Report Key7969344
Report Number1416980-2018-06565
Device Sequence Number1572549
Product Code FPA
UDI-Device Identifier00085412004648
UDI-Public(01)00085412004648
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K060074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/17/2023
Device Catalogue Number2N3326
Device Lot NumberUR18F17068
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/19/2018
Supplement Date Received by Manufacturer11/05/2018
Initial Report FDA Received Date10/16/2018
Supplement Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-