| Model Number |
8888541028P |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, when removing the protective bandage before starting dialysis, the atrial catheter shaft disconnected from the (push-in) fixed y-piece.Only a quick disconnection of the residual catheter prevented a more serious complication (e.G.Air embolism/blood loss.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, when removing the protective bandage before starting dialysis, the atrial catheter shaft disconnected from the (push-in) fixed y-piece.Only a quick disconnection of the residual catheter prevented a more serious complication (e.G.Air embolism/blood loss.It was stated that there was no blood loss.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection noted: that the blue port, red port, clamps, silicone extensions and bifurcate was received.The cannula of the catheter was cut, disengaged and not received.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the cut and disengaged cannula may occur when mishandled during clinical application.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|