| Catalog Number |
PXT261412 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Thrombus (2101)
|
| Date of Event |
11/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional devices involved in this event:pxc141400/(b)(4) and pxc181200/(b)(4).According to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and/or require intervention include, but are not limited to infection, fistula, occlusion, and pseudoaneurysm.
|
| |
|
Event or Problem Description
|
|
On an unknown date in 2009, a patient underwent endovascular surgery using gore® excluder® aaa endoprostheses.On an unknown date in 2012, the patient presented to the hospital due to deep thrombosis venous which were reportedly managed with anticoagulants.On an unknown date in 2012, the patient reportedly presented with weakness, as well as pain in the leg and back.Imaging reportedly identified a retroperitoneal collection of blood.It was reported through a retroperitoneal access the clots were collected.It was noted that the origin of the bleeding is suspected to be a pseudo aneurysm.It was reported the pseudo aneurysm was treated by an iliac ligature.An axilo-femoral bypass was also reportedly unsuccessfully performed.The patient then reportedly underwent an amputation of the left leg leaving 10cm of the patient¿s femur.It was reported the patient developed a cutaneous fistula.The patient reportedly had colonization of bacteria due to the fistula as the incision is above the retroperitoneal access.On an unknown date in 2017, imaging showed a cutaneous fistula and a suspected type iii endoleak.
|
| |
|
Additional Manufacturer Narrative
|
|
Added h6: code 213 the review of the manufacturing and sterilization evaluation paperwork verified that these lots met all pre-release specifications.Images from one time point available for evaluation: post-implantation cta dated (b)(6) 2017 were returned to gore and an imaging evaluation was performed.According to the images summary: the length from the left renal artery to the proximal circumferential device appears to be 22.3mm.There does not appear to be flow in the contralateral portion of the device that extends into the lci.There appears to be air-like densities within the aneurysmal sac.There appears to be air-like densities within the contralateral portion of device on the left.The aortic borders at the level of the mid-distal device(s) appear to be unclear.There appears to be a surgical clip at the level of the left femoral head.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected: date of event.
|
| |
|