Catalog Number 151630510
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 09/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the insert was removed and replaced due to infection.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot 8640945 device history review manufactured on 13 oct 2017.16 parts were manufactured per specification and all raw materials met specification.Parts sterilized on 23 oct 2017, sterilization cycle met the validated parameters (certificate 61367).There were no ncs or deviations.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR RP INSRT SZ 5 10MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key7973113
Report Number1818910-2018-72704
Device Sequence Number1681864
Product Code NJL
UDI-Device Identifier10603295047780
UDI-Public10603295047780
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151630510
Device Lot Number8640945
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/25/2018
Supplement Date Received by Manufacturer11/23/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date11/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-