MAUDE Adverse Event Report: CORDIS CASHEL PRECISE PRO RX US CAROTID SYST; SELF EXPANDABLE STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL PRECISE PRO RX US CAROTID SYST; SELF EXPANDABLE STENT Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult or Delayed Positioning (1157)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing telephone and fax numbers of cordis cashel are respectively, (b)(4).During an unknown procedure.An 8x30 precise pro rx us carotid system did not cross the lesion.There was no patient injury.No other information was reported.The device was returned for analysis.A non-sterile precise pro rx us carotid system was received for analysis.Per visual inspection, the stent was partially deployed.No other damages or anomalies were detected.Per dimensional analysis, the outer diameter (od) measurements were taken at three different places and were found within specification.Functional test could not be performed due to the nature of the event.A product history record (phr) review of lot 17656553 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The phr review does not suggest that the event reported by the customer could be related to the manufacturing process.The event reported by the customer as ¿stent delivery system - failure to cross" could not be confirmed due to the nature of the event and the relationship to the device.The dimensional analysis results were found within specification.Analysis of the returned device indicated the stent was partially deployed.The event identified by fal analysis as ¿stent delivery system (sds)-ses deployment difficulty - premature deployment¿ was determined via visual analysis.Per the instructions for use, which is not intended as a mitigation, ¿if resistance is met during delivery system introduction, the system should be withdrawn, and another system should be used.If any resistance is met during delivery system withdrawal, advance the outer sheath until the marker contacts the catheter tip and withdraw the system as one unit.The exact cause of the events could not be conclusive determined.However, procedural and handling factors such as the application of excessive force, as well as vessel characteristics, while unknown, may have contributed to the reported events.Per the instructions for use, which is not intended as a mitigation, ¿precise® pro rx nitinol stent system is shipped with the tuohy borst valve in the open position.Care should be taken not to pre-deploy the stent.The device should be prepped in the tray.After prep, extract the stent delivery system from the tray.Examine the device for any damage.Evaluate the distal end of the catheter to ensure that the stent is contained within the outer sheath.Do not use if the stent is partially deployed.If a gap between the catheter tip and outer sheath tip exists, open the tuohy borst valve and gently pull the inner shaft in a proximal direction until the gap is closed.Lock the tuohy borst valve after the adjustment by rotating the proximal valve end in a clockwise direction.¿ neither the product analysis nor the phr review suggests that the event could be related to the manufacturing process; therefore, no corrective action will be taken at this time.
 
Event or Problem Description
An 8x30 precise rx self expanding stent did not.Cross the lesion.There was no patient injury.Analysis of the returned device indicated the stent was partially deployed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISE PRO RX US CAROTID SYST
Common Device Name
SELF EXPANDABLE STENT
Manufacturer (Section D)
CORDIS CASHEL
cahir road
co. tipperary
cashel
EI 
MDR Report Key7973237
Report Number9616099-2018-02447
Device Sequence Number16564472
Product Code NIM
UDI-Device Identifier20705032036488
UDI-Public20705032036488
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Model NumberN/A
Device Catalogue NumberPC0830RXC
Device Lot Number17656553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2018
Initial Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-