Batch review performed on 15 october 2018.Lot 170259: (b)(4) items manufactured and released on 08 june 2017.Expiration date: 25.05.2022 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Gmk-sphere tibial tray fixed cemented size t3-i4 r reference 02.12.T3i4r (k121416).Lot 171841: (b)(4) items manufactured and released on 12 july 2017.Expiration date: 2022-06-29.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medical affairs director on october 17, 2018: one year after primary cemented tkr, a revision is performed because the surgeon decided the original implant positions to be suboptimal and potential cause for patient dissatisfaction.According to a very approximated analysis of the postoperative radiographs (approximation due to low quality of the images), the components seem to have been placed substantially as planned, except posterior tibial slope (lower than planned) and femoral flexum (higher than planned).We cannot determine the influence that these variations may have had on the final outcome nor the reasons that led the surgeon to make these intraoperative changes, but we find no reason to suspect a malfunctioning device.Investigation performed by patient match department on september 26, 2018: our analysis of the myknee process of this case found no deviations from the standard procedures.Each step has been performed correctly.
|