DEPUY MITEK LLC US RAPIDLOC PLA /PLA 2/0 PAN 12; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 228321
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/22/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The lot number is currently unavailable; therefore, the exp date is unavailable.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This is report 2 of 4 for the same event.It was reported by the affiliate in (b)(6) that during an unspecified surgical procedure, it was observed that while using rapidloc anchor, the surgeon was only successful in one of them and complained that the thread was very fragile since it broke before finishing the suture adjustment.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.Product complaint # (b)(4).Investigation summary: the device is not available for a physical evaluation.This complaint is not confirmed.Without physically evaluating the device, a definite root cause cannot be determined.Furthermore, no lot number was supplied which precludes conducting a dhr review or a lot specific search in the complaints handling system.At this point in time, no corrective action is required, and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.: udi: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate associated medwatch report numbers: 1221934-2018-54968, 1221934-2018-54966, 1221934-2018-54969.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDLOC PLA /PLA 2/0 PAN 12
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7973322
Report Number1221934-2018-54967
Device Sequence Number1487813
Product Code MAI
UDI-Device Identifier10886705010226
UDI-Public10886705010226
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number228321
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/24/2018
Supplement Date Received by Manufacturer12/01/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date12/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-