• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VERSALOK DEPLOYMENT GUN *EA; ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VERSALOK DEPLOYMENT GUN *EA; ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR Back to Search Results
Catalog Number 214710
Medical Device Problem Codes Positioning Failure (1158); Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The lot number is currently unavailable.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported by the sales rep via phone that the suture anchor did not deploy from their versaloc gun during a rotator cuff repair.There were no patient consequences or delays.The case was completed with a competitors device.No new bone holes needed to be created.The device is being returned.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and inspected.The complaint cannot be confirmed.An inserter was placed inside the deployment shaft, and the gun was fired.It was observed that the inserter was capable of pushing forward which would then lead to the anchor deploying.The device appears to be in good condition, with the exception of some minor signs of wear which would not impede deployment.A device history record (dhr) review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7973404
Report Number1221934-2018-54970
Device Sequence Number16457450
Product Code LXH
UDI-Device Identifier10886705004416
UDI-Public10886705004416
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214710
Device Lot Number1003001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Initial Date Received by Manufacturer 11/06/2017
Supplement Date Received by Manufacturer10/25/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date10/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-