| Catalog Number |
PXB35-06-57-080 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
09/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The visi-pro balloon-expandable biliary stent system from the implant procedure was not received for evaluation.No cine images from either the implant, post/follow-up, or explant procedure were received for evaluation.From the explant procedure a portion of the visi-pro balloon-expandable stent was received entangled within the gold/tungsten hoop of a snare.Based on the radiopaque marker hoops and stent strut pattern, the returned stent segment is from a visi-pro stent.The returned stent segment is approximately 25mm in length.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The patient had an x-ray for an unrelated issue; it was observed on the x-ray that a previously implanted visi-pro stent in the left upper extremity venous system was fractured, procedure completed by dr (b)(6) ((b)(6) 2018).The visi-pro had been implanted to treat a fibrous lesion which was pre-dilated using a 4x80 balloon.The visi-pro was implanted in the left proximal to treat a fibrous lesion, artery diameter 6mm the lesion was pre-dilated using a 4x80 balloon.The x-ray showed a portion of the stent had migrated to the left pulmonary artery.The physician removed the fractured stent using a snare, pulling the fractured portion into the sheath and removing as a unit by dr (b)(6) ((b)(6) 2018).No injury to the patient reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|