MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number PRD-SC30-125
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.(b)(4).
 
Event or Problem Description
During a peripheral atherectomy procedure using a csi orbital atherectomy device (oad), a device malfunction occurred.During preparation for the procedure, it was noted that the saline solution would not advance through the tubing of the peripheral device.The procedure was completed with balloon angioplasty and there were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
The reported oad was received for analysis.When the device was tested for functionality, there was no fluid flow observed through the saline line.Examination of the saline line revealed an adhesive occlusion in the saline hose barb connector on the hose junction.There were no other issues observed with the device.At the conclusion of the device analysis investigation, the reported event was confirmed.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
MDR Report Key7973603
Report Number3004742232-2018-00308
Device Sequence Number1773755
Product Code MCW
UDI-Device Identifier10852528005039
UDI-Public(01)10852528005039(17)191031(10)173425
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K110389
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/30/2019
Device Model NumberPRD-SC30-125
Device Catalogue Number70056-03
Device Lot Number173425
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer10/17/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Congenital Anomaly;
Patient Age73 YR
-
-