| Model Number |
PRD-SC30-125 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.(b)(4).
|
| |
|
Event or Problem Description
|
|
During a peripheral atherectomy procedure using a csi orbital atherectomy device (oad), a device malfunction occurred.During preparation for the procedure, it was noted that the saline solution would not advance through the tubing of the peripheral device.The procedure was completed with balloon angioplasty and there were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported oad was received for analysis.When the device was tested for functionality, there was no fluid flow observed through the saline line.Examination of the saline line revealed an adhesive occlusion in the saline hose barb connector on the hose junction.There were no other issues observed with the device.At the conclusion of the device analysis investigation, the reported event was confirmed.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
|
| |
|