| Catalog Number |
1200015 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Hernia (2240)
|
| Date of Event |
11/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No conclusion can be made.While the clinician has stated the issue to be "possible device related", based on the information provided we are unable to determine to what extent, if any the phasix st mesh may have caused or contributed to the patient's hernia recurrence.Recurrence is a known inherent risk of hernia repair surgery.Regarding recurrence the warning section of the instructions-for-use state, "to prevent recurrences when repairing hernias, phasix st mesh must be large enough to provide sufficient overlap beyond the margins of the repair/primary closure.Careful attention to mesh fixation placement and spacing will help prevent excessive tension or gap formation between the mesh and fascial tissue.A review of the manufacturing records was performed and found that the lot was manufactured to specification.If additional information is provided, a supplemental emdr will be submitted.Remains implanted.
|
| |
|
Event or Problem Description
|
|
It was reported to davol that a patient who is part of the phasix st study (b)(4), experienced a hernia recurrence.(b)(6) 2016 - the subject patient underwent a robotic assisted umbilical hernia repair with implant of the phasix st mesh.The technique was performed intra-abdominally.The phasix st mesh is reported to have been positioned so that its edges extend beyond the margins of the defect by at least 5cm.The mesh was fixed with absorbable monofilament suture.(b)(6) 2018 -the subject patient underwent a ct scan in which a hernia recurrence was noted.The hernia recurrence is assessed to be within 5cm of the index hernia repair.The hernia recurrence has been assessed by the clinician as moderate in severity, possibly related to the study device and not related to the index procedure.No intervention has been reported at this time, condition is listed as ongoing.
|
| |
|
Event or Problem Description
|
|
It was reported to davol that a patient who is part of the phasix st study dvl-he-017, experienced a hernia recurrence.(b)(6) 2016 - the subject patient underwent a robotic assisted umbilical hernia repair with implant of the phasix st mesh.The technique was performed intra-abdominally.The phasix st mesh is reported to have been positioned so that its edges extend beyond the margins of the defect by at least 5cm.The mesh was fixed with absorbable monofilament suture.(b)(6) 2018 -the subject patient underwent a ct scan in which a hernia recurrence was noted.The hernia recurrence is assessed to be within 5cm of the index hernia repair.The hernia recurrence has been assessed by the clinician as moderate in severity, possibly related to the study device and not related to the index procedure.No intervention has been reported at this time, condition is listed as ongoing.Addendum: it was reported that on 11/19/2018 the subject patient underwent additional surgery to repair the previously diagnosed recurrent hernia.During this procedure the phasix st mesh was explanted.Per clinician the phasix st mesh had a portion of which did not incorporate.
|
| |
|
Additional Manufacturer Narrative
|
|
As reported during the explant procedure it was noted that a portion of the mesh had not incorporate, it is unclear if this contributed to the hernia recurrence.The information provided does not indicate what may have caused a portion of the mesh to not incorporate.There is no change to the initial conclusion, the results remain inconclusive.If additional information is provided, a supplemental emdr will be submitted.Section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|