• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL SHUNT KIT BURR HOLE, 16 MM, MEDIUM PRESSURE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL SHUNT KIT BURR HOLE, 16 MM, MEDIUM PRESSURE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 9040 E
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that upon opening the implant, it was found that the peritoneal end of the shunt was broken.It was noted there was no patient involved in the event.
 
Additional Manufacturer Narrative
The shunt end of the peritoneal catheter was damaged/broken.No other unusual characteristics were noted.It is unknown how or when this damage occurred.The instructions for use (ifu) that accompanies the device cautions that, ¿low tear strength is a characteristic of most unreinforced silicone elastomer materials.Care must be taken with the handling and placement of the silicone elastomer catheter tubing to avoid cuts, nicks or tears.¿ all catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL SHUNT KIT BURR HOLE, 16 MM, MEDIUM PRESSURE
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7973768
Report Number2021898-2018-00502
Device Sequence Number16564630
Product Code JXG
UDI-Device Identifier00643169466340
UDI-Public00643169466340
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K831678
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model Number9040 E
Device Catalogue Number9040 E
Device Lot NumberE36075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2018
Supplement Date Received by Manufacturer11/30/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date12/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-