Model Number ETBF3216C145E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Occlusion (1984)
Date of Event 10/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 50mm diameter abdominal aortic aneurysm.It was reported that the patient presented emergently two days later and it was reported that during the index procedure the celiac branches were mis-identified as the renal arteries and as a result the stent graft was placed across both renal arteries and sma.It was reported that while the stent graft was higher than anticipated the stent graft did exclude the aneurysm.The patient was placed on dialysis.As the mesenteric artery was being perfused by a patent celiac artery it was decided that additional treatment was not required.Per the physician, the cause of the event was due to user error.No additional clinical sequelae were reported and the patient is being monitored.
 
Additional Manufacturer Narrative
Fdc code added at investigation completion.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7974008
Report Number2953200-2018-01478
Device Sequence Number16447700
Product Code MIH
UDI-Device Identifier00613994990983
UDI-Public00613994990983
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2020
Device Model NumberETBF3216C145E
Device Catalogue NumberETBF3216C145E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/07/2018
Supplement Date Received by Manufacturer12/20/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date01/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-