Model Number XOM UNKNOWN BUR |
Device Problems
Material Separation (1562); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis found the broken bur is stuck in the nose cone.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The device was returned with no information.There was no patient involvement.
|
|
Manufacturer Narrative
|
Visually, the bur was broken into at least two pieces.The break point was consistent with being bent until broken and there were tool marks present.There was excess wear to the tang.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from service and repair that the drill head was broken and noting that only the bur base half remained stuck in the drill.
|
|
Search Alerts/Recalls
|