| Model Number |
16186555 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for investigation as product remains in-situ.A review of x-ray film provided confirms the reported event.Labeling review: ".Potential risks identified with the use of this system, which may require additional surgery, include:· bending, fracture or loosening of implant component(s) loss of fixation.".".The lock screws on the multi-load lock screw starter must be compressed by the central sleeve after delivery of each lock screw to ensure proper engagement into the tulip.".".Step 8: reline mod tulip head attachment ensure there is no soft tissue on or around the shank that may inhibit the polyaxial head from being inserted properly.".
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018 patient underwent a spinal procedure.On (b)(6) 2018 during a post-operative follow-up visit it was noted the tulip head on the l5 screw shank came off.No patient harm or injury reported.
|
| |
|