| Catalog Number |
357.515 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.The investigation could not be completed; no conclusion could be drawn.As no product was received.Device was used for treatment, not diagnosis.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2018, the ball hex screwdriver split down from the handle during sterile processing.The handle had been repaired by the customer¿s shop department with a non synthes handle.There was no patient involvement.This complaint involves one (1) device.This report is 1 of 1 of for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A product investigation was conducted.Visual inspection: the ball hex screwdriver 8 mm for ti femoral nails (part number 357.515, lot number unknown) was received at us customer quality.Upon visual inspection, the shaft was identified as subcomponent of ball hex screwdriver 8 mm for ti femoral nails (part number 357.515).The original synthes wooden handle and the dowel pin was not returned.It also appears that the customer had welded the shaft onto a non-synthes handle.There remaining of the shaft showed surface wear, consistent with device use.The visual inspection of the device revealed that the device components, handle and the dowel pin are missing and were not returned for investigation.The complaint condition is confirmed.Document/specification review: the relevant drawing(s) was reviewed; a dhr review could not be performed as the lot numbers is unknown.Dimensional inspection: dimensional analysis was completed on the shaft component of screwdriver.The diameter of the shaft measured and is within specification based on relevant drawing.The diameter of the ball hex measured and is within specification based on relevant drawing.All dimensional inspections completed with calipers.The length of the shaft (outside of the non-synthes handle) could not be performed accurately as the device has been altered.Dimensional analysis was not performed on the non-synthes handle.Conclusion: based on visual inspection, it can be confirmed that the original handle and dowel pin are no longer present and that the device has been modified.Further investigation cannot be completed as the original device has been modified.No design or manufacturing defect or deficiency was observed during the investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.
|
| |
|