| Model Number |
TV-IL1422140-J |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Occlusion (1984); Claudication (2550)
|
| Date of Event |
09/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device remains implanted in the patient and will not be returned for evaluation.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
An ovation ix abdominal stent graft system was implanted to treat an abdominal aortic aneurysm.It was discovered during a routine follow-up ct that the patient has a right side ovation limb that is occluded.The patient reported experiencing claudation at the follow-up visit.Patient is currently stable.Reportedly, there are plans to treat the occlusion with thrombectomy/embolectomy and kissing stents.No additional event or patient information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging.Requests were made and denied response were received.As such, event determination, off label conditions, related patient harms and patient disposition could not be independently assessed.The review of manufacturing lot confirmed all devices met specifications prior to release.No additional investigations of this reported complaint are planned, endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.
|
| |
|
Event or Problem Description
|
|
Subsequent to the initial report, additional information was provided reporting that re-intervention was performed with a non-endologix device.
|
| |
|