(b)(4).Device evaluation by manufacturer: field service engineering (fse) visited the customer to address the reported event.During servicing, fse confirmed no pressure on the instrument during power up.Fse found an open drain valve, closed it, and then brought the instrument into stand-by status.Fse then ran a few patients but the pressure was still at 0 mpa.The observed retention time was at 0.54 minutes.Fse tried adjusting flow rate, but the retention times did not rise.Fse replaced the purge and uptake valves and then set the flow rate on the pump to 1.00 ml/min.The retention time then increased to 0.57 minutes.Fse readjusted the flow rate to 1.04 ml/min with patient results coming in at retention 0.59 minutes.The analyzer was loaded with the current software version and quality controls were run with acceptable ranges.Patient samples were run without any issue.The instrument was operating as expected.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 05-sept-2017 through aware date (b)(4) 2018.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 6.3: error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.101 pressure low: the pressure will not rise because the pump is unable to run due to air bubbles in the pump check valve.If the elution buffer is empty, place a new elution buffer and execute reagent change.Next, execute drain flush.See "chapter 5 section 5.5: pump air removal".Execute manual pumping using the pump key in the main screen (second screen), and open and close the drain valve 2 or 3 times.If the pressure rises when the drain valve is closed, the operation is complete.If the pressure still does not rise or stabilize, execute drain flush again.In addition, confirm that the drain valve is securely closed.The most probable cause of the reported event was due to failure of the purge and uptake valves.
|
A customer reported that the pressure was reading 0 mpa on the g8 instrument.When the issue occurred, the customer changed the filter, powered the instrument off and on, checked for leaks, and performed a drain flush three times.Technical support (ts) instructed the customer to disconnect the line going into the filter housing and turn on the pump.The pump did not power on.The instrument was down.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|