• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTXSFT0103
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 80.0 and 121.0 cm from the proximal end.The embolization coil was intact with its pusher assembly and had offset coil winds.Resistance was encountered while attempting to advance the smart coil through the hub of the demonstration microcatheter and the smart coil could not be advanced any further.Conclusions: evaluation of the first returned smart coil revealed that the embolization coil winds were offset.If the introducer sheath is not properly aligned within the hub of the microcatheter prior to advancement, damage such as this may occur.During the functional test, resistance was encountered as the embolization began to take shape inside the hub of the demonstration microcatheter and the smart coil could not be advanced any further.The offset coil winds likely contributed to the smart coil not being able to be advanced out of its introducer sheath and into the non-penumbra microcatheter during the procedure.Evaluation of the second returned smart coil revealed that the embolization coil winds were offset.If the introducer sheath is not properly aligned within the hub of the microcatheter prior to advancement, damage such as this may occur.During the functional test, resistance was encountered while attempting to load the introducer sheath over the pusher assembly kink and the sheath could not be advanced any further.The offset coil winds likely contributed to the smart coil not being able to be advanced out of its introducer sheath and into the non-penumbra microcatheter during the procedure.The neuron max and non-penumbra microcatheter identified in the complaint were not returned to penumbra for evaluation.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-02047.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the middle cerebral artery (mca) using penumbra smart coils (smart coils).During the procedure, the physician placed three smart coils in the target vessel using a neuron max 6f 088 long sheath (neuron max) and a non-penumbra microcatheter.Next, the physician opened a new smart coil but it would not advance out of its introducer and into the microcatheter.After two failed attempts, the smart coil was removed.The physician then opened another smart coil but the same issue occurred; therefore, it was removed.The procedure was completed using a two additional smart coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7975428
Report Number3005168196-2018-02046
Device Sequence Number1682481
Product Code HCG
UDI-Device Identifier00814548015194
UDI-Public00814548015194
Combination Product (Y/N)Y
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2021
Device Catalogue Number400SMTXSFT0103
Device Lot NumberF70245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2018
Initial Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-