• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD; INC. JELCO® CATHLON® I.V. CATHETERS; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD; INC. JELCO® CATHLON® I.V. CATHETERS; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Catalog Number 446611
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Report source: (b)(6).
 
Event or Problem Description
Information was received indicating that while the nurse was attempting to pull out a smiths medical jelco® cathlon® i.V.Catheter, the plastic part broke off in the patient.It was reported that approximately 2mm of the plastic came out.Subsequently, the patient had the severed catheter removed in interventional radiology.There were no further reported adverse patient effects.
 
Additional Manufacturer Narrative
Device evaluation summary: one jelco cathlon iv catheter assembly was returned for evaluation.Immediate inspection of the sample showed the catheter had been cut off at approximately 5.9mm from the hub nose.Further examination of the cut showed the surface was jagged and highly damaged.During production, to assure that the products meet functional requirements the dimensions of the catheter components (eyelet, tubing and hub) are tightly controlled.Once assembled, these products are 100% inspected to demonstrate the catheter tube is properly secured to the hub.In addition, catheter samples from each lot number are subjected to destructive testing to determine minimum break strength.Based on the investigation, the complaint allegation was confirmed.User interface is noted as the problem source, due to evidence that the damage of the device was consistent with the use of the catheter against the device ifu (instructions for use); either through a needle reinsertion or use of a lancet or scissor near the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JELCO® CATHLON® I.V. CATHETERS
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
SMITHS MEDICAL ASD; INC.
6000 nathan lane n
minneapolis, MN 55442
MDR Report Key7975531
Report Number3012307300-2018-04092
Device Sequence Number1254978
Product Code FOZ
UDI-Device Identifier15019315118548
UDI-Public15019315118548
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K990236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number446611
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer01/23/2019
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date02/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-