| Model Number |
AAB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to johnson and johnson vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the monofocal intraocular lens (model aab00, +22.0 diopter) was explanted from patient¿s right eye because of patient experiencing blurry vision.There was no incision enlargement, no vitrectomy performed during this procedure.The replacement lens used was a model zxr00 +24.5 diopter.Patient was doing good post exchange.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes.Returned to manufacturer on: 10/10/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under 12x microscope magnification was performed by a qualified final inspection manufacturing specialist.There were pits at the posterior side of the optic area.This type of reject is acceptable as per hb0017 ¿cosmetic specification of acrylic one-piece lenses¿.No other cosmetic abnormalities found on the lens.Lens was further measured as per tm4759 rev.01 - dimensional measurement of returned complaint sample, and all measurements (haptic width, optic diameter, overall diameter and haptic thickness) were within specification.Complaint event is not confirmed.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|