MAUDE Adverse Event Report: STRYKER GMBH BONE SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L48MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH BONE SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L48MM; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 657448S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
When inserting a screw, the thread stripped from screw.
 
Event or Problem Description
When inserting a screw, the thread stripped from screw.
 
Additional Manufacturer Narrative
The reported event that bone screw variax2 t10, full thread, 3.5mm / l48mm was alleged of 'device stripped/twisted' could be confirmed only looking at the picture though.Based on the received picture, the root cause was attributed to be user related.The failure was caused by possible inadequate angulations during screw insertion.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.Looking at the picture it revealed the following: that some metal was getting coiled up, possibly right after the first turn during screw insertion.The surgeon stopped approximately 3/4 of the screw length as he must have seen the metal piece hanging off.The length of the metal indicates that the screw was positioned on an unfavorable angle so that the material was getting coiled up.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L48MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key7976366
Report Number0008031020-2018-00875
Device Sequence Number13525192
Product Code HRS
UDI-Device Identifier07613327074253
UDI-Public07613327074253
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K132502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Catalogue Number657448S
Device Lot NumberV50302
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/20/2018
Supplement Date Received by Manufacturer11/20/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age24 YR
-
-