• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Model Number G52264
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, an advance 35 lp low profile balloon catheter ruptured between 4 and 5 atmospheres (atm) due to heavy calcification in the distal superficial femoral artery (sfa).Slight angulation of the vessel was reported.Visipaque 320 contrast was used in a 50/50 mix with saline.The angioplasty procedure was successfully completed with another device of the same type.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation ¿ evaluation.A review of the dimensional verification, complaint history, device history record, instructions for use (ifu), manufactures instructions, quality control, trends, and a visual inspection of the returned device was conducted during the investigation.The visual inspection of the returned device confirmed that the catheter was smooth, clean, and not damaged.The length of the balloon catheter measured within specification.A 2.1cm longitudinal rupture was noted in the balloon.No other nonconformances were noted with the device.Due to the rupture accurate measurements of the balloon length and diameter could not be obtained.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, a review of the manufactures instructions and quality control procedures was conducted, and no gaps were discovered.Moreover, an ifu is provided with the device, which states ¿choose a balloon appropriate to lesion length and vessel diameter.Upon removal from package, inspect the catheter to ensure no damage has occurred during shipping." how supplied: "store in a dark, dry cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred." it goes on to say, ¿if resistance is met while advancing the balloon dilation catheter, determine the cause and proceed with caution.¿ based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be traced to the device, but may lie with the patient¿s condition has heavy calcification was noted.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7976607
Report Number1820334-2018-03108
Device Sequence Number1484099
Product Code DQY
UDI-Device Identifier10827002522647
UDI-Public(01)10827002522647(17)210412(10)8768811
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/12/2021
Device Model NumberG52264
Device Catalogue NumberPTA5-35-135-6-2.0
Device Lot Number8768811
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2018
Initial Date Received by Manufacturer 10/03/2018
Supplement Date Received by Manufacturer12/27/2018
Initial Report FDA Received Date10/17/2018
Supplement Report FDA Received Date01/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age51 YR
Patient Weight111
-
-