MAUDE Adverse Event Report: VASCUTEK LTD GELWEAVE; GELWEAVE STRAIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VASCUTEK LTD GELWEAVE; GELWEAVE STRAIGHT Back to Search Results
Catalog Number 731508
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 09/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A review of the retained manufacturing and quality records (with specific emphasis of the base material permeability and device porosity testing) confirmed that the product was manufactured to the intended specifications.The whole batch of 32 passed porosity testing well below acceptance limit of 26ml/min with minimum reading of 0ml/min and highest reading of 5ml/min.A review of similar events was performed for all woven devices that have been reported to have leaked.The incident rate for reported leakage events was confirmed as 0.017% (complaints v sales).The review included all similar events reported to vascutek between jan2013 - sep 2018 inclusive.No negative trend was identified during the review.The device was misused as the supplied ifu states: "the prosthesis must be immersed in a sterile saline solution of 5 minutes.Failure to rinse for 5 minutes could lead to the graft being more susceptible to leakage when implanted." further action is not planned; however, the issue will be tracked and trended as part of the on-going complaints trending and reporting process and if an adverse trend develops action may be taken at that time.Vascutek consider this case closed.
 
Event or Problem Description
The reported event was reported as follows: blood bled through during implant.Hospital does not pre-soak grafts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GELWEAVE
Common Device Name
GELWEAVE STRAIGHT
Manufacturer (Section D)
VASCUTEK LTD
newmains avenue
inchinnan business park
renfrewshire, PA4 9 RR
UK  PA4 9RR
MDR Report Key7976917
Report Number9612515-2018-00019
Device Sequence Number1755907
Product Code DSY
UDI-Device Identifier05037881105079
UDI-Public05037881105079
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K952293
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/18/2018
Report Date (Section F) 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Catalogue Number731508
Device Lot Number17320285 1217
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date09/20/2018
Device Age4 MO
Event Location Hospital
Date Report to Manufacturer09/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2018
Initial Report FDA Received Date10/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
-
-