| Model Number |
MMT-523NAP |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had compromised forced system and no delivery alarm.The customer¿s blood glucose level was unknown at the time of the incident.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Unable to confirm compromised force sensor system alarm and unable to perform displacement test, basic occlusion test, occlusion test due to reservoir tube lip broken off.The test reservoir does not stay locked in place due to reservoir tube lip broken off.Unable to perform prime test and excessive no delivery test due to loose or protruded drive support disk.
|
| |
|
Search Alerts/Recalls
|