• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-55-user had an inaccurate reference: competitor's meter: the end user is comparing results obtained from trividia's bgm system to the results from a competitor's bgm system.(b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 291, 140, 133 and 151 mg/dl.Customer also stated he compared truemetrix air meter to two other meters at the doctor's office during a routine scheduled appointment.Customer stated truemetrix air read 38 points higher; actual meter to meter comparison results were not provided.The customer's expected fasting blood glucose test result range is 100 - 120 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 09/30/2018 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1:114mg/dl date: (b)(6) 2018 time:9:31pm fasting, result 2:291mg/dl date: (b)(6) 2018 time:9:52am fasting, result 3:140mg/dl date: (b)(6) 2018 time:8:11am fasting, result 4:133mg/dl date: (b)(6) 2018 time:6:49am fasting, result 5:151mg/dl date: (b)(6) 2018 time:4:23pm fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7977734
Report Number1000113657-2018-01061
Device Sequence Number19170657
Product Code NBW
UDI-Device Identifier00052569139042
UDI-Public(01)00052569139042
Combination Product (Y/N)N
Initial Reporter StateKY
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2018
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMU2422
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2018
Initial Report FDA Received Date10/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-