| Model Number |
G34979 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Abscess (1690); Unspecified Infection (1930)
|
| Date of Event |
09/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device lot history record indicated the device was manufactured to specifications.A total of 10 devices was produced from the lot.A review of the cbi complaint database did not reveal any additional complaints involving the reported lot number.Per the ifu, the device is supplied sterile to the end user.The device should be used "with caution in infection regions." infection is noted amongst the list of potential complications.The root cause, of the abscess and infection, is not likely device related.Such an occurrence could be procedure related and/or acquired post operatively.Cook biotech incorporated has a validated sterilization cycle with a sterility assurance level of (sal) 10^-6.The validated sterility cycle and all parameters of all devices are verified prior to release for distribution.
|
| |
|
Event or Problem Description
|
|
Professor (b)(6) performed an endoscopic transnasal repair of a skull base fracture in the anterior ethmoid roof.The (b)(6) 2018 procedure included implanting a biodesign duraplasty graft in the (b)(6) female patient.Prior to surgery the patient presented with a spontaneous csf leak and had a known history of generalized seizures.The patient had an uneventful recovery and was seen post repair regularly.On (b)(6) 2018, the patient presented with a generalized grand-mal seizure.An urgent brain ct showed a possible small intracranial abscess at the graft site.The patient was admitted to the icu and underwent emergency surgery to explant the graft.The patient was in the hospital for 7 days on antibiotics.The patient recovered and was discharged.
|
| |
|
Search Alerts/Recalls
|