| Model Number |
E2350H-GNSB |
| Medical Device Problem Codes |
Fire (1245); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Full thickness (Third Degree) Burn (2696)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the pencil caught fire.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic partial cecectomy, the raytec in the surgical field caught fire which resulted from prolonged use of the device at the trocar site to gain coagulation due to lack of effectiveness from the pencil with no change in settings.The pencil did not catch fire.The raytec fire was immediately extinguished.The patient sustained a burn from the heat accumulation at trocar site.The burn was full thickness skin necrosis, 0.5 cm of necrotic tissue.The injury care included sharp debridement and packing of the wound.Patient is currently still dressing the wound, but appears to be healing.The associated generator in use was a forcetriad generator.
|
| |
|