| Model Number |
3888-45 |
| Medical Device Problem Codes |
Device Damaged Prior to Receipt by User (1570); Human-Device Interface Problem (2949); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from the manufacturer's representative regarding a patient who was implanted with a neurostimulator for spinal cord stimulation.It was reported that the physician went to implant lead and electrode was protruding and stylet was not holding in place.It was a new lead straight out of the box.Lead to be sent back for analysis.New lead was used.Issue resolved at this time.No further complications reported/anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis results were not available at the time of this report.A follow up report will be sent once analysis is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the lead (lot# 3888-45) identified that the butt joint of the outer insulation was separated at the proximal end of the lead.This was determined to be use error.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
A correction was made to reflect most accurate evaluation code- conclusion.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|