• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. HYDROSOFT HELICAL 10 VTA; EMBOLIZATION COIL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. HYDROSOFT HELICAL 10 VTA; EMBOLIZATION COIL Back to Search Results
Model Number 100206H2HS-V
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device returned to the manufacturer was found to have the implant coil attached to the pusher, which was inconsistent with the reported details.It was determined that the returned device was not associated with the reported complaint.The actual device used for the procedure was not returned for analysis; therefore, the root cause cannot be determined.The instructions for use (ifu) identifies premature or difficult coil detachment as potential complications associated with use of the device.
 
Event or Problem Description
It was reported that during re-positioning of the coil in the aneurysm, the coil detached without use of the controller.The coil remains implanted in the aneurysm.There was no reported intervention or patient injury as a result of the issue.The patient was reported to be okay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROSOFT HELICAL 10 VTA
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key7979388
Report Number2032493-2018-00219
Device Sequence Number19115600
Product Code HCG
UDI-Device Identifier00816777022899
UDI-Public(01)00816777022899(11)180425(17)230331(10)1804251W4
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K070656
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number100206H2HS-V
Device Lot Number1804251W4
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2018
Initial Report FDA Received Date10/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient Weight73
-
-