Boston scientific received information that the remote home monitoring system issues an alert for high out of range impedance measurements of greater than 2000 ohms for the cardiac resynchronization therapy pacemaker (crt-p) and another manufacturer¿s right atrial (ra) lead where the lead safety switch (lss) was triggered.
Review of the data found one higher impedance reading of 1400 ohms from four months earlier.
Review of the logbook found inappropriately stored atrial tachy response (atr) episodes from noise due to unipolar configuration after the lss was triggered.
Further review found one inappropriate atr due to oversensing of noise from the minute ventilation signal eight months earlier.
The field representative noted that the rate response trend had already been programmed off in the crt-p as the patient is pacemaker dependent.
Boston scientific technical services (ts) discussed possible causes of abrupt impedance changes.
The field representative that she performed maneuvers and could not create noise and all lead measurement were good.
Ts discussed the programming options and noted that the other manufacturer¿s rv lead currently is displaying stable impedance measurements.
The crt-p and ra lead remain in service and no adverse patient effects were reported.
This device was included in the recent mv sensor signal oversensing advisory population.
|